Galileo Gold
Περιγραφή
Αποστολή σε Φίλο σας
Reviews
Manufacturer: | Hamilton Medical AG | ||
Ventilator Type: | Positive Pressure Invasive Conventional Nitric Oxide |
||
Ventilator Class: | Intensive Care Unit (ICU) Transport CPAP |
||
Patient Population Range: | Neonatal Pediatric Adult |
||
Description | |||
Features: | Adaptive Support Ventilation (ASV) maintains an operator-preset, minimal Minute Ventilation independent of the patient's activity. | ||
Characteristics | |||
Modes--Invasive: | Volume-controlled modes --- ASSIST/CONTROL; SIMV (PSV); MMV; SPONT; (S)CMV; ASV Pressure-controlled modes --- ASSIST/CONTROL; SIMV (PSV); CPAP; APV |
||
Waveforms: | Sine, square, decelerating, 50% decelerating | ||
Operating Range of Controls: | Special functions: Manual breath, inspiratory/expiratory hold, nebulizer, 100% O2, sigh, apnea backup; CMV rate: 5 to 120 b/min; SIMV rate: 1 to 60 b/min; Tidal volume: 10 to 2000 ml; PEEP/CPAP 0 to 50 cmH2O; Oxygen: 21 to 100%; I:E ratio: 1:9 to 4:1; Inspiratory time: 10 to 80% of cycle time (0.1 to 3 s); Pause time 0 to 70% of cycle time (0 to 8 s); Peak flow 1 to 180 l/min; Pressure trigger 0.5 to 10 cmH2O below PEEP/CPAP; Flow trigger 0.5 to 15 l/min; Base flow 4 to 30 l/min, depending on flow trigger setting; Pressure control 5 to 100 cmH2O; Pressure support 0 to 100 cmH2O; Pressure ramp 25 to 200 ms TO..63; Expiratory trigger sensitivity 10 to 40% of inspiratory peak flow; % minute volume (ASV) 10 to 350%; Bodyweight (ASV) 10 to 200 kg |
||
Triggering: | Pressure trigger: 0.5 to 10 cm H2O below PEEP/CPAP. Flow trigger: 0.5 to 15 l/min. | ||
Size: | 17.5 W x 24.5 D x 61 H | ||
Weight: | 97 lbs. | ||
Other Characteristics: | Monitoring: Real-time waveforms -- volume, flow, airway pressure, auxiliary pressure; Pressures -- peak, mean, minimum, plateau, PEEP/CPAP, autoPEEP, P(01); Flow -- inspiratory peak, expiratory peak; Volume -- tidal, minute, leakage; Time -- inspiratory, expiratory, I:E ratio, total frequency, spontaneous frequency; Others -- Inspiratory resistance, expiratory resistance, static compliance, inspiratory time constant, expiratory time constant, imposed work of breathing, pressure time product, rapid shallow breathing index, delivered oxygen concentration; Options: Nebulizer, communications interface including RS-232C port, remote nurse’s call, I:E signal |
||
Characteristics (Non - Technical) | |||
Approvals: | IEC 601-1/EN 60601-1, IEC 601-1-2/EN 60601-1-2, EN 794-1, ISO 7767. CE. CSA, FDA | ||
Specifications | |||
Power Requirements: | Electrical requirements: 100 to 240 V ~ +/-10%, 50/60 Hz; O2 and air supply: 200 to 600 kP a (29 to 86 psi) |
||
Alarms: | Low/high minute volume: 0 to 50 l/min; Low/high pressure: 0 to 110 cmH2O; Low/high tidal volume: 0 to 3000 ml; Low/high rate: 0 to 130 b/min; Apnea time: 10 to 60 s; Air trapping: 1 to 10 l/min; Special: Oxygen concentration, disconnection, oxygen/air supply, loss of PEEP, check settings, flow sensor alarms, power supply, inoperative |
||
Mounting Options: | Standard trolley or shelf mount | ||
Other Specifications: | Operating environment: Temperature – 10 to 40 C, (-10 to 60 C during transport); Humidity -- 30 to 75% noncondensing, (5 to 85% noncondensing during transport); Altitude – 3000 m; Backup battery time: 20 minutes minimum (fully charged) |
||
Monitored Variables : | |||
Tidal Volume | 10 to 2000 ml |